Syringe pump Infutek 405
 
Syringe pump Infutek 400
 
 
 
 
 




















FAIL (the browser should render some flash content, not this).

Syringe pump Infutek 400

Simtek - Infutek 400 Syringe Infusion Pump is indispensable wherever high-potency drugs are to be precisely delivered at controlled flow rates over an extended period.
This is a basic equipment for : Intensive care units (ICU) , operating theatres, emergency, anaesthesiology, cardiology, neonatology, paediatrics, oncology and general hospital wards. Ideally suited for administration of anti-cancer drugs, oxytocics, chemotherapy medication, anaesthesia & life saving drugs.

ONE PUMP - MULTIPLE OPTIONS
User freindly & Safe
  • Accepts 10,20,30 and 50mL syringe
  • 0.1 mL/Hr to 450mL/Hr flow rate
  • Programmable to accept any brand of syringe
  • Target Volume setting
  • Target time setting
  • Elapsed time display
  • All information about infusion status on
  • LCD Screen without pressing any key
  • Display of user selected Drug
  • Infused Volume display
  • Built in rechargeble battery
  • Automatic syringe size detection
  • Bolus / Purge facility
  • Upgradable to PCA
  • Variable occlusion pressure
  • Easy to use with LED displays together with LCD displays.
  • Programmable KVO(keep vein open)
SYRINGE:
10, 20, 30, 50/60 ML
ACCEPTABLE SYRINGES
All available standard indian
(Dispovan,B.D.etc)and imported
syringes(can be programmed for
other syringes as per
customer`s requirements.)
Custom Programmable
Programmable to any brand Accuracy Volume : +/- 2% , Time : +/- 2%
Syringe. with Automatic syringe size
detection.
Flow Rate Range
0.1-120ml/Hr (10ml Syringe)
0.1-200ml/Hr (20ml Syringe)
0.1-300ml/Hr (30ml Syringe)
0.1-450ml/Hr (50ml Syringe)
Infused Volume Range
0.1 - 999.9ml
Bolus Rate
300ml/hr for 10ml Syringe
500ml/hr for 20ml Syringe
650ml/hr for 30ml Syringe
999ml/hr for 50ml Syringe
Bolus Volume
Programmable by user
Bolus Interval
A minimum interval of time can
be preset between two Bolus
dose to avoid accidental over
infusion, 1min to 24 Hrs.
OCCLUSION
Low, Medium, High Levels for all sizes of syringes
ALARMS:
Occlusion, End of syringe, Low Battery,
Syringe displaced, No Syringe, Wrong
Syringe, Pre Alarm for END, Invalid
rate, target Volume, Target Time.
PCA:
(optional) Max. volume and Interval
can be predefined by user.
ACCURACY:
Volume : +/- 2% , Time : +/- 2%
VOLUME LIMITS:
From .1ml to capacity of the syringe.
ELAPSED TIME:
Visual display of elapsed time
POWER:
240 VAC, +/- 15 %, 50/60 Hz 12VA
BATTERY TYPE
Ni Mh , 12V, 1600mAHr
BATTERY BACK-UP
2-3 Hrs ( Typ.) @ 200 ml/Hr.
4-5 Hrs (Typ.) @ 5 ml/Hr.
BATTERY RECHARGING
20 Hrs. For full Charging
WEIGHT
2.7 kg
DIMENSION
240(w) X 145(d) X 140 (h)
STANDARDS
Tested as per IEC 60601.1
   
2008 © Copyright SIMTEK. All rights reserved.  |  Powered by : Yoko9.com